Nexera XS inert - Specs

Ultra High Performance Liquid Chromatograph

Add Bookmark

Allowable maximum pressure 105 MPa
Internal rinsing Standard equipment (3 dil)
Features Nexera XS inert combines excellent resistance to corrosion with the inertness of surfaces in contact with the sample and mobile phases. The system provides excellent chromatographic separation performances for a wide range of applications.

 

  Model Name Features
System controller SCL-40*
CBM-40*
CBM-40lite*
Color screen with touch panel and system controller with reservoir tray
Number of connectable units
: 8 (Using option: 12) , 4(CBM-lite, Excluding built-in solvent delivery unit)
Connectable unit : Solvent delivery unit: max. 4, Autosampler: 1, Column oven: max. 4, Detector: max 2, etc.
Solvent Delivery Pump LC-40D XSi Pumping method : Parallel-type double plunger (approx. 10 μL/1 stroke)
Allowable maximum pressure : 105 MPa
Flow rate settings range : 0.0001 to 3.0000 mL/min (1.0 to 105 MPa), 3.0001 to 5.0000 mL/min (1.0 to 80 MPa), 5.0001 to 10.0000 mL/min (1.0 to 22 MPa)
Flow rate accuracy :≦± 1% (under specified conditions)
Gradient mode : High-pressure gradient (2 or 3 solvents) Quaternary low-pressure gradient
Gradient Mixer MR Mixer(Inert) MR 20 µL 
MR 180 µL 
MR 40 µL LPGE 
MR 300 µL LPGE
 
Degassing Unit DGU-403*
DGU-405*
Number of degassed solvents : 3(DGU-403), 5(DGU-405)
Degassed low line capacity : 400 μL/1 line
Operating temperature range : 4 to 35°C
Autosampler SIL-40C XSi Injection method : Total-volume Injection (standard), loop injection (optional)
Allowable maximum pressure : 105 MPa
Injection volume : 0.1 to 50 μL, 0.1 to 2000 μL (optional)
Injection volume accuracy : ≦± 1% (5 μL injection, n = 20)
Injection cycle time : ≦6.7 seconds (under specified conditions)
Column Oven CTO-40S Temperature control type : Forced air circulation
Cooling Method : Electronic cooling
Temperature control range :
Room temperature -10 to 85 ℃(CTO-40S)
Room temperature -10 to 100 ℃(CTO-40C)
Temperature accuracy :±0.8℃(CTO-40S),  ±0.5℃(CTO-40C)
Temperature precision : 0.2 ℃(±0.1 ℃、CTO-40S), 0.1 ℃(±0.05 ℃、CTO-40C)
UV-VIS Detector SPD-M40*
Please contact your local Shimadzu representative.
Light source : Deuterium (D2) lamp
Number of diode elements : 1024
Wavelength range : 190 to 700 nm
Wavelength accuracy : ≦± 1 nm
Wavelength reproducibility : ≦± 0.1 nm
Flow Cell UHPLC inert cell Cell volume: 8 μL, Cell path length: 10 mm
Low-diffusion inert cell Cell volume: 2.5 μL, Cell path length: 5 mm
Inert cell Cell volume: 12 μL, Cell path length: 10 mm
pH monitor pHM-40 Measurement range:pH 1 to 14
Measurement error:pH ± 0.1
Ambient operating temperature:4 to 35 ℃
Usable humidity range:20 to 85 %
Mobile phase monitor    
Software Labsolutions LC/GC Multi LC-PDA, LabSolutions LCMS
Please contact your local Shimadzu representative.


 

Nexera XS inert

Dual pump

-

{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":4636,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":4636,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}