Nexera series - Configuration
Ultra High Performance Liquid Chromatograph
- Overview
- Features
- Applications
- Specs
- Downloads
- Configuration
- Columns
- Consumables
- Options
- Case Studies
- Support
System Controller
SCL-40 | CBM-40 | CBM-40lite | |
---|---|---|---|
Monitor | Touch panel LabSolutions™ Web monitor |
LabSolutions Web monitor |
LabSolutions Web monitor |
Connectable unit |
Solvent delivery unit: max. 4, Autosampler: 1, Column oven: max. 4, Detector: max 2, etc. | ||
Number of connectable units | 8 (Using option: 12) | 4 (Excluding built-in solvent delivery unit) | |
Event input/output | Input: 1, output: 2 | ||
Analog board |
Up to two channels (option) | Up to one channel (option) | ̶ |
Communication | Ethernet | ||
Reservoir tray | Built-in | ̶ | |
Dimensions, weight | W 260×D 500×H 140 mm、6 kg | W 260×D 500×H 72 mm、5 kg | ̶ |
Operating temperature range | 4 to 35°C | ||
Power supply | AC 100–240 V, 50 VA, 50/60 Hz | Supplied from solvent delivery unit |
Solvent Delivery Pump
LC-40D |
LC-40D XR LC-40B XR |
LC-40D XS LC-40D XSi |
LC-40D X3 LC-40B X3 |
|
---|---|---|---|---|
Pumping method | Parallel-type double plunger (approx. 10 μL/1 stroke) | |||
Allowable maximum pressure | 44 MPa | 70 MPa | 105 MPa | 130 MPa |
Flow rate settings range | 0.0001-5.0000 mL/min (1.0-44 MPa) 5.0001-10.0000 mL/min (1.0-22 MPa) |
0.0001-3.0000 mL/min (1.0-70 MPa) 3.0001-5.0000 mL/min (1.0-44 MPa) 5.0001-10.0000 mL/min (1.0-22 MPa) |
0.0001-3.0000 mL/min (1.0-105 MPa) 3.0001-5.0000 mL/min (1.0-80 MPa) 5.0001-10.0000 mL/min (1.0-22 MPa) |
0.0001-3.0000 mL/min (1.0-130 MPa) 3.0001-5.0000 mL/min (1.0-80 MPa) 5.0001-10.0000 mL/min (1.0-22 MPa) |
Flow rate accuracy | ≤± 1% or ± 2 μL/min, whichever greater (under specified conditions) |
≤± 1% (under specified conditions) | ||
Flow rate accuracy | ≤ 0.06% RSD or 0.02 minSD, whichever greater | |||
Gradient mode |
High-pressure gradient (2 or 3 solvents) Quaternary low-pressure gradient | High-pressure gradient (2 solvents (LC-40B XR standard) or 3 solvents) Quaternary low-pressure gradient (Only available for LC-40D XR) | High-pressure gradient (2 or 3 solvents) Quaternary low-pressure gradient | High-pressure gradient (2 solvents (LC-40B X3 standard) or 3 solvents) Quaternary low-pressure gradient (Only available for LC-40D X3) |
Gradient range of set concentrations | 0 to 100% (0.1% step) | |||
Gradient range of set concentrations | ± 0.5% (under specified conditions) | |||
Wetted materials | SUS316L, Hastelloy® C, PEEK, PTFE, Sapphire, Ruby | SUS316L, Hastelloy C, PEEK, PE, Sapphire, Ruby Titanium、Titanium alloys、Nickel alloys、PEEK、PTFE、Ruby、Sapphire、Hastelloy C、PE (LC-40D XSi) |
||
Available pH range | 1 to 14 | |||
Automatic rinsing kit | Option | Standard equipment | ||
Degassing unit | 1 unit connectable | LC-40D XR:1 unit connectable LC-40B XR: 2 units connectable | 1 unit connectable |
LC-40D X3:
1 unit connectable LC-40B X3: pre-installed (5 port builtin), 1 unit connectable |
Dimensions | W 260×D 500×H 140 mm | LC-40D X3: W 260×D 500×H 140 mm LC-40B X3: W 260×D 500×H 210 mm |
||
Weight | 10 kg | LC-40D XR:10 kg LC-40B XR:13 kg |
12 kg | LC-40D X3:12 kg LC-40B X3: 21 kg |
Operating temperature range | 4 to 35°C | |||
Power supply | AC 100-240 V、50/60 Hz | |||
150 VA | LC-40D XR:150VA LC-40B XR:180 VA |
150 VA | LC-40D X3:150 VA LC-40B X3:180 VA |
Autosampler
SIL-40 SIL-40C |
SIL-40 XR SIL-40C XR |
SIL-40C XS SIL-40C XSi |
SIL-40C X3 | ||
Injection method |
Total-volume Injection (standard), loop injection (optional, except SIL-40C XSi) | ||||
Allowable
maximum pressure
|
44 MPa | 80 MPa | 105 MPa | 130 MPa | |
Injection volume | 0.01-100 μL | 0.01-50 μL | |||
0.01 to 2000 μL (optional, except SIL-40C XSi) | |||||
Injection volume
accuracy
|
≤ ± 1% (5 μL injection, n = 20) | ||||
Linearity | ≥ 0.9999 | ||||
Injection cycle time | ≤ 6.7 seconds (under specified conditions) | ||||
Samples for processing | 288( microtiter plate, 96 well × 3 plates) 1152( microtiter plate, 384 well × 3 plates) 252( 1 mL sample vial, 84 × 3 plates) 162( 1.5 mL sample vial, 54 × 3 plates) 84( 4 mL sample vial, 28 × 3 plates) 36( 10 mL sample vial, 12 × 3 plates) 72( 1.5 mL micro tube, 24 × 3 plates)) |
||||
Injection volume reproducibility |
RSD≦1.0% (0.5-0.9 μL) RSD≦0.5% (1.0-1.9 μL) RSD≦0.25%( 2.0-4.9 μL) RSD≦0.15%( More than 5.0 μL) RSD<0.5% (typically, 0.5 μL, except SIL-40C XSi) RSD<0.25%( typically, 1.0 μL, except SIL-40C XSi) |
||||
Carryover |
≤ 0.0025% (without rinse) ≤0.0005% (with rinse, typically) (under specified conditions) |
≤ 0.0015%(without rinse) ≤ 0.0003%(with rinse, typically)) (under specified conditions) |
|||
Dip rinsing outside the needle and injection port rinsing |
Standard equipment | ||||
Pumping rinse
outside the needle
|
Option | Standard equipment | |||
Internal rinsing (3 dil) | Option | Option(SIL-40C XS) Standard equipment (SIL-40C XSi) |
Standard equipment | ||
Sample cooler |
SIL-40:None SIL-40C:Standard equipment (Air-circulation temperature control type) |
SIL-40 XR:None SIL-40C XR:Standard equipment (Air-circulation temperature control type) |
Standard equipment (Air-circulation temperature control type) |
||
Sample cooler temperature setting range |
4 to 45°C
(Room temperature needs to be less than 30°C and humidity needs to be less than 70% to set 4°C)
|
||||
Sample cooler temperature accuracy |
±2 ℃(sensor position ± 0.5°C) | ||||
Wetted material | SUS316L, DLC, PEEK, GFP, PTFE, FEP, ETFE, sapphire, ceramics, PPS, FFKM PEEK、ceramics、PTFE、ETFE、FEP、GFP、sapphire、PPS、FFKM (SIL-40C XSi) |
||||
Available pH range | 1 to 14 | ||||
Dimensions | W 260×D 500×H 280 mm(SIL-40C/40C XR/40C XS/40C XSi/40C X3Protrusion adds 140 mm to the depth) | ||||
Weight | SIL-40:17 kg SIL-40C:24 kg |
SIL-40 XR:17 kg SIL-40C XR:24 kg |
24 kg | ||
Operating
temperature range
|
4 to 35 ℃ | ||||
Power supply | Cooler model | AC 100-240 V, 50/60 Hz | |||
Non cooler model | SIL-40:150 VA SIL-40C:400 VA |
SIL-40 XR:150 VA SIL-40C XR:400 VA |
ー
|
Column Oven
CTO-40C | CTO-40S | |
---|---|---|
Temperature control type | Forced air circulation | |
Cooling Method | Electronic cooling | |
Temperature control range | Room temperature − 10°C to 100°C | Room temperature − 10°C to 85°C |
Temperature accuracy | ±0.5 ℃ | ±0.8 ℃ |
Temperature precision | 0.1 ℃(±0.05 ℃) | 0.2 ℃(±0.1 ℃) |
Containable column size and number |
Up to 250 mm L. column × 6 or 300 mm L. column × 3 |
Up to 100 mm L. column × 6 or 300 mm L. column × 3 |
Dimensions, weight | W 260×D 500×H 415 mm、21 kg | W 130×D 500×H 553 mm、15 kg |
Operating temperature range | 4 to 35°C | |
Power supply | AC 100~120 V/AC 220-240 V(Automatic switching) 400 VA、50/60 Hz |
AC 100-240 V、300 VA、50/60 Hz |
SPD-40 | SPD-40V | |
---|---|---|
Light source | Deuterium (D2) lamp | Deuterium (D2) lamp, tungsten lamp |
Wavelength range | 190 to 700 nm | 190 to 1000 nm |
Bandwidth | 8 nm | |
Wavelength accuracy | ≤ ± 1 nm±1 nm | |
Wavelength reproducibility | ≤ ± 0.1 nm | |
Drift | ≤ 0.1 × 10−3 of AU/h (under specified conditions) | |
Noise | 1 Wavelength mode: ≤ 4.0 × 10−6 AU, 2 Wavelength mode: ≤ 10.0 × 10−6 AU (under specified conditions) | |
Linearity | 2.5 AU (under specified conditions) | |
Standard flow cell | Optical path length: 10 mm, Cell volume: 12 μL, Pressure: 12 MPa Material of wetted parts: SUS316L, PFA, quartz, PEEK |
|
Cell temperature control range | 19 to 50°C, 1°C Step | |
Optional flow cell |
UHPLC cell (optical path length: 10 mm, cell volume: 8 μL, equipped with temperature control function) Semi-micro cell (optical path length: 5 mm, cell volume: 2.5 μL, equipped with temperature control function) Conventional cell (optical path length: 10 mm, cell volume: 12 μL, equipped with temperature control function) Inert cell (optical path length: 10 mm, cell volume: 12 μL, equipped with temperature control function) UHPLC inert cell (optical path length:10 mm、cell volume:8 μL、equipped with temperature control function) Low-diffusion inert cell (optical path length:5 mm、cell volume:2.5 μL、quipped with temperature control function) Preparative cell (optical path length: 0.1/0.2/0.5 mm, cell volume: 0.8/1.6/4.0 μL) Micro flow cell (optical path length: 3 mm, cell volume: 0.21 μL) Maximum pressure cell (optical path length: 10 mm, cell volume: 12 μL) |
|
Available pH range | 1 to 13 (Cell quartz might be damaged by a mobile phase of pH >10.) | |
Dimensions , weight | W 260 × D 500 × H 140 mm, 11 kg | |
Operating temperature range | 4 to 35°C | |
Power supply | AC 100–240 V, 150 VA, 50/60 Hz |
Photodiode Array Detector
SPD-M40 | ||
---|---|---|
Light source | Deuterium (D2) lamp, Tungsten lamp | |
Number of diode elements | 1024 | |
Wavelength range | 190 to 800 nm | |
Wavelength accuracy | ≤ ±1 nm | |
Wavelength reproducibility | ≤ ±0.1 nm | |
Slit width | 1.2 nm、8 nm | |
Spectral resolution | ≤ ± 1.4 nm | |
Drift | ≤ 0.4×10-3 AU/h (under specified conditions) | |
Noise | ≤ 4.5 × 10−6 AU (under specified conditions) | |
Linearity | 2.5 AU (under specified conditions) | |
Standard flow cell | Optical path length: 10 mm, Cell volume: 12 μL, Pressure: 12 MPa Material of wetted parts: SUS316L, PFA, quartz, PEEK |
|
Cell temperature control range | 19 to 50°C, 1°C Step | |
Optional flow cell |
UHPLC cell (optical path length: 10 mm, cell volume: 8 μL, equipped with temperature control function) Semi-micro cell (optical path length: 5 mm, cell volume: 2.5 μL, equipped with temperature control function) Conventional cell (optical path length: 10 mm, cell volume: 12 μL, equipped with temperature control function) Inert cell (optical path length: 10 mm, cell volume: 12 μL, equipped with temperature control function) UHPLC inert cell (optical path length:10 mm、cell volume:8 μL、equipped with temperature control function) Low-diffusion inert cell (optical path length:5 mm、cell volume:2.5 μL、quipped with temperature control function) Preparative cell (optical path length: 0.1/0.2/0.5 mm, cell volume: 0.8/1.6/4.0 μL, equipped) Micro flow cell (optical path length: 3 mm, cell volume: 0.21 μL) Maximum pressure cell (optical path length: 10 mm, cell volume: 12 μL) |
|
Available pH range | 1 to 13 (Cell quartz might be damaged by a mobile phase pH >10.) | |
Dimensions , weight | W 260 × D 500 × H 140 mm, 10 kg | |
Operating temperature range | 4 to 35°C | |
Power supply | AC 100–240 V, 180 VA, 50/60 Hz |
Capillary cell type Photodiode Array Detector
SPD-M30A | ||
---|---|---|
Light source | Deuterium (D2) lamp | |
Number of diode elements | 1024 | |
Wavelength range | 190 to 700 nm | |
Wavelength accuracy | ≤ ±1 nm | |
Wavelength reproducibility | ≤ ±0.1 nm | |
Slit width | 1 nm, 8 nm | |
Spectral resolution | ≤ 1.4 nm | |
Drift | ≤ 4.0 × 10−6 AU (under specified conditions) | |
Noise | ≤ 4.0 × 10−6 AU (under specified conditions) | |
Linearity | 2.0 AU (under specified conditions) | |
Cell | Standard cell: Optical path length: 10 mm, Capacity: 1 μL, Pressure: 8 MPa Optional high-sensitivity cell: Optical path length: 85 mm, Capacity: 9 μL, Pressure: 8 MPa Optional low-dispersion cell PLUS: Optical path length: 10 mm, Capacity: 1 µL, Pressure: 8 MPa |
|
Dimensions, weight | W 260×D 500×H 140 mm,12 kg | |
Operating temperature range | 4 to 35°C | |
Power supply | AC 100–240 V, 150 VA, 50/60 Hz |
Spectrofluorometric Detector
RF-20A | RF-20AXS | ||
---|---|---|---|
Light source |
Xenon lamp | Xenon lamp Low-pressure mercury lamp (to check wavelength accuracy) |
|
Wavelength range | 200 to 650 nm | 200 to 750 nm | |
Spectral bandwidth | 20 nm | ||
Wavelength accuracy |
±2 nm |
||
Wavelength precision
|
±0.2 nm | ||
S/N |
Water Raman peak S/N ≥ 200 Low background S/N ≥ 9000 |
Water Raman peak S/N ≥ 2000 Low background S/N ≥ 12000 |
|
Range of cell temperature control | ̶ | Room temperature − 10°C to 40°C, 1°C step | |
S/N
|
Water Raman peak S/N ≥1200 Low background S/N ≥ 9000 |
Water Raman peak S/N ≥ 2000 Low background S/N ≥ 12000 |
|
Cell | Standard conventional cell: volume 12 μL, maximum pressure 2 MPa Optional semi-micro cell: volume 3 μL, maximum pressure 2 MPa Option inert cell: volume 12 µL, maximum pressure 2 MPa |
||
Function | Simultaneous measurement of four wavelengths, Wavelength scanning | ||
Dimensions | W 260×D 500×H 210 mm | ||
Weight | 17 kg | 18 kg | |
Operating temperature range | 4 to 35°C | ||
Power supply | AC 100–240 V, 400 VA, 50/60 Hz |
Differential Refractive Index Detector
RID-20A | |
---|---|
Measurement range | 1 to 1.75 RIU |
Noise | ≤ 2.5×10-9 RIU |
Drift | ≤ 1×10-7 RIU/h |
Range | A mode: 0.01 × 10−6 to 500 × 10−6 RIU P, L-mode: 1 × 10−6 to 5000 × 10−6 RIU |
Response
|
0.05 to 10 sec, 10 steps |
Polarity – Change | Available |
Zero adjustment | Auto zero, Optical zero, Fine zero |
Maximum flow rate
|
20 mL/min (150 mL/min in option)
|
Range of cell temperature control | 30 to 60°C |
Cell | Volume 9 μL, Maximum pressure 2 MPa (equipment : 0.4 MPa) |
Dimensions, weight | W 260 × D 420 × H 140 mm, 12 kg |
Operating temperature range | 4 to 35°C |
Power supply | AC 100–240 V, 150 VA, 50/60 Hz |
Conductivity Detector
CDD-10Avp | |
---|---|
Cell volume | 0.25 μL |
Cell constant | 25 μS·cm⁻¹ |
Material of wetted parts | PEEK, SUS316 |
Maximum use pressure | 2.9 MPa(30 kgf/cm2) |
Response
|
0.05 to 10 s, 10 steps |
Zero adjustment | Auto-zero function, Baseline-shifting function |
Dimensions, weight | W 260 × D 420 × H 140 mm, 6 kg |
Operating temperature range
|
4 to 35°C |
Power supply | AC 100–240 V, 250 VA, 50/60 Hz |
Evaporative Light-Scattering Detector
ELSD-LT II | ELSD-LT III | |
---|---|---|
Nebulizing method | Siphon Splitting | Siphon Splitting |
Light source | LED | Semiconductor laser |
Detector | Photomultiplier | Photodiode |
Temperature Setting Range | Room temperature + 5°C to 80°C | Room temperature to 100 °C |
Nebulizer Gas
|
Nitrogen or air* | Nitrogen or air* |
Gas flow rate, gas pressure | Up to 3.0 mL/min, up to 450 kPa | Up to 3.0 mL/min, up to 450 kPa |
Mobile Phase Flow Rate (Standard Nebulizer) |
0.2 to 2.5 mL/min | 0.2 to 2 mL/min |
Analog output
|
0 to 1 V | 0 to 1 V |
Dimensions | W 260×D 550×H 450 mm | W 250×D 530×H 330 mm |
weight | 20 kg | 15.5 kg |
Operating temperature range | 5 to 40°C | 4 to 35°C |
Operation humidity range | ≤ 80% (Room temperature + 5°C to 31°C), ≤ 50% (Room temperature + 31°C to 40 °C) |
≦80% (4~31℃), ≦50% (31~35℃) |
Power supply | AC 100 V, 210 VA, 50/60 Hz | 100 to 240 VAC, 120 VA, 50/60 Hz |
* Supply gas at a pressure of about 350 kPa.
An air compressor may also be used.
A filter (P/N: 228-45528-92) is also available for filtering out moisture and other matter from the compressor.
For more details, contact a Shimadzu sales representative.
- Monitor mobile phase levels remotely from a PC or smartphone
- Avoid running out of mobile phase or rinse solution mid-analysis
- Fits a range of bottle sizes from 1 to 5L
Description | ||
---|---|---|
Mobile Phase Monitor (Controller) | This is a controller to connect the bottle holders. It can connect two reserve trays of bottle holders. |
|
Mobile Phase Monitor (1L bottle holder) | One holder can accommodate two 1L bottles. | |
Mobile Phase Monitor (Large bottle holder) | One holder can accommodate one large bottle. |
※ The large bottle holder can hold bottles up to a diameter of approx. 180 mm.
※ To change between the two holder types, a field engineer will need to carry out the modifications.
Versatile layout can accommodate different bottle combinations
Two trays of bottles can be monitored from one MPM controller.
- Large bottle holder ×1
- 1L bottle holder ×1
- 1L bottle holders ×3
- Large bottle holders ×2
ANALYTICAL INTELLIGENCE
Automated support functions utilizing digital technology, such as M2M, IoT, and Artificial Intelligence (AI), that enable higher productivity and maximum reliability.
pH monitor
pHM-40 | |
Available pH range | 1 to 14 |
Precision | pH ±0.1 (under specified conditions) |
Drift | pH ±0.1 /10 h (under specified conditions) |
Maximum flow rate | 10 mL/min |
Allowable maximum pressure | 0.1 MPa |
Cell volume | approx. 80 μL |
Available liquid temperature | 4 to 60 °C |
Wetted materials | glass, PEEK, PCTFE, silicon |
Calibration point | 1 to 5 point is settable. |
Dimensions [mm], weight | W 130 × D 393 × H 206, 6 kg |
Power supply | DC 5 V, 5 VA |
Plate Changer
PLATE CHANGER | ||
---|---|---|
Samples for processing
(includes two plates of autosampler)
|
1 PLATE CHANGER
1536 (microtiter plate, 96 well × 16 plates) 864 (deep-well plate, 96 well × 9 plates)
6144 (microtiter plate, 384 well × 16 plates) 3456 (deep-well plate, 384 well × 9 plates)
756 (1 mL sample vial, 84 × 9 plates) 486 (1.5 mL sample vial, 54 × 9 plates)
252 (4 mL sample vial, 28 × 9 plates) 108 (10 mL sample vial, 12 × 9 plates)
|
3 PLATE CHANGERS
4224 (microtiter plate, 96 well × 44 plates)
2208 (deep-well plate, 96 well × 23 plates)
16896 (microtiter plate, 384 well × 44 plates) 8832 (deep-well plate, 384 well × 23 plates)
1932 (1 mL sample vial, 84 × 23 plates) 1242 (1.5 mL sample vial, 54 × 23 plates)
644 (4 mL sample vial, 28 × 23 plates) 276 (10 mL sample vial, 12 × 23 plates)
|
Sample cooler temperature setting range |
Air-circulation temperature control type, 4 to 45°C (Room temperature needs to be less than 30°C and humidity needs to be less than 70% to set 4°C) |
|
Dimensions, weight | W 170 × D 500 × H 560 mm (Protrusion adds 140 mm to the depth), 26 kg | |
Operating temperature range | 4 to 35°C | |
Power supply | AC 100−240 V, 400 VA, 50/60 Hz |
High-Pressure Gradient Using Dual Pump, Photodiode Array Detector
P/N | Description | Q'ty |
---|---|---|
228-65500-46/58 | SCL-40 | 1 |
228-65078-46/58 | LC-40B X3 | 1 |
228-65120-46/58 | SIL-40C X3 | 1 |
228-65201-46/58 | CTO-40S | 1 |
228-65302-46/58 | SPD-M40 | 1 |
228-64724-41 | UHPLC Cell | 1 |
228-72652-42 | Mixer MR 40 μL (with a mixer recognition device) |
1 |
228-70254-42 | Piping Kit B, for High-Pressure GE ID0.1 | 1 |
228-70247-43 | Nexera Series Cable Installation Kit, C | 1 |
Power Cord, Length: 2.5 m | 5 | |
228-57647-43 | Hand Tool Kit | 1 |
High-Pressure Gradient Using Dual Pump, Plate Changer, Photodiode Array Detector
P/N | Description | Q'ty |
---|---|---|
228-65500-46/58 | SCL-40 | 1 |
228-65078-46/58 | LC-40B X3 | 1 |
228-65120-46/58 | SIL-40C X3 | 1 |
228-65110-46/58 | Plate Changer | 1 |
228-65201-46/58 | CTO-40S | 1 |
228-65302-46/58 | SPD-M40 | 1 |
228-64724-41 | UHPLC Cell | 1 |
228-72652-42 | Mixer MR 40 μL (with a mixer recognition device) |
1 |
228-70254-42 | Piping Kit B, for High-Pressure GE ID0.1 | 1 |
228-70247-43 | Nexera Series Cable Installation Kit, C | 1 |
Power Cord, Length: 2.5 m | 6 | |
228-57647-43 | Hand Tool Kit | 1 |
High-Pressure Gradient Using Two Pumps, Photodiode Array Detector
P/N | Description | Q'ty |
---|---|---|
228-65500-46/58 | SCL-40 | 1 |
228-65076-46/58 | LC-40D X3 | 2 |
228-65019-58 | DGU-405 | 1 |
228-65120-46/58 | SIL-40C X3 | 1 |
228-65202-46/58 | CTO-40C | 1 |
228-65302-46/58 | SPD-M40 | 1 |
228-64724-41 | UHPLC Cell | 1 |
228-72652-42 | Mixer MR 40 μL (with a mixer recognition device) |
1 |
228-70254-42 | Piping Kit B, for High-Pressure GE ID0.1 | 1 |
228-70247-43 | Nexera Series Cable Installation Kit, C | 1 |
Power Cord, Length: 2.5 m | 6 | |
228-57647-43 | Hand Tool Kit | 1 |
High-Pressure Gradient Using Dual Pump, UV-VIS Detector
P/N | Description | Q'ty |
---|---|---|
228-65500-46/58 | SCL-40 | 1 |
228-65078-46/58 | LC-40B X3 | 1 |
228-65120-46/58 | SIL-40C X3 | 1 |
228-65202-46/58 | CTO-40C | 1 |
228-65313-46/58 | SPD-40V | 1 |
228-64724-42 | UHPLC Cell | 1 |
228-72652-42 | Mixer MR 40 μL (with a mixer recognition device) |
1 |
228-70254-42 | Piping Kit B, for High-Pressure GE ID0.1 | 1 |
228-70247-44 | Nexera Series Cable Installation Kit, D | 1 |
Power Cord, Length: 2.5 m | 5 | |
228-57647-43 | Hand Tool Kit | 1 |
Low-Pressure Gradient, Photodiode Array Detector
P/N | Description | Q'ty |
---|---|---|
228-65500-46/58 | SCL-40 | 1 |
228-65076-46/58 | LC-40D X3 | 1 |
228-65016-46/58 | Low-Pressure Gradient Kit | 1 |
228-65019-58 | DGU-405 | 1 |
228-65120-46/58 | SIL-40C X3 | 1 |
228-65201-46/58 | CTO-40S | 1 |
228-65302-46/58 | SPD-M40 | 1 |
228-64724-41 | UHPLC Cell | 1 |
228-65020-41 | Mixer Installation Kit, MR 40 μL LPGE (with a mixer recognition device) | 1 |
228-70247-42 | Nexera Series Cable Installation Kit, B | 1 |
Power Cord, Length: 2.5 m | 5 | |
228-57647-43 | Hand Tool Kit | 1 |
228-36573 | 1/16"NYLON COUPLING | 2 |
228-56687-42 | SF NUT,1/8 | 4 |
228-56687-43 | SF FERRULE,1/8 | 4 |
High Performance Liquid Chromatography Nexera™ lite, ELSD-LT III
P/N | Description | Q'ty |
---|---|---|
LC-40D | 2 | |
CBM-40 lite | 1 | |
DGU-403 | 1 | |
SIL-40C | 1 | |
VIAL PLAT 1.5mL | 2 | |
CTO-40C | 1 | |
ELSD-LT III | 1 | |
RESERVOIR TRAY | 1 | |
MR 180μL II | 1 | |
PIPING KIT A, for HPGE ID0.3 | 1 | |
PEEK Male Nut | 2 | |
PEEK Tube, 0.13mmIDx3000mm | 1 | |
REGULATOR,W/FILTER | 1 | |
Optical Fiber Cable Assy, Plastic, for Nexera Series | 1 | |
CORD,UC-975-N01 | 5 | |
TOOL KIT 1 | 1 | |
LabSolutions LC Workstation Single | 1 | |
LC LabSolutions | 1 | |
SHIMADZU LabTotal™ for LC 1.5 mL, Glass A | 1 | |
Shodex HILICpak VG-50 4E(250 × 4.6 mm I.D., 5 μm) | 1 |
Nexera™ XR High Performance Liquid Chromatograph
P/N | Description | Q'ty |
---|---|---|
LC-40B XR | 1 | |
SCL-40 | 1 | |
DGU-403 | 1 | |
SIL-40CXR | 1 | |
VIAL PLAT 1.5mL | 2 | |
CTO-40C | 1 | |
SPD-M40 | 1 | |
MR 180μL II | 1 | |
PIPING KIT B, for HPGE ID0.1 | 1 | |
Optical Fiber Cable Assy, Plastic, for Nexera Series, C 1 CORD,UC-975-N01 | 5 | |
POWER OUTLET UNIT | 1 | |
TOOL KIT | 1 | |
Glass Reservoir Bottle, w/Cap, 5pcs | 1 | |
Mobile Phase Monitor (Contoller) | 1 | |
Mobile Phase Monitor(bottle holder) | 2 | |
Mobile Phase Monitor AC Adapter | 1 | |
CMD CABLE | 1 | |
MPM Bottle Holder Kit,Eeathquake-R | 1 | |
BELT | 1 | |
LOCKING PLATE | 1 | |
SYSTEM BASE KIT B | 2 | |
LabSolutions LCGC,DB Multi LC-PDA | 1 | |
LC LabSolutions | 1 |
High Performance Liquid Chromatography
P/N | Description | Q'ty |
---|---|---|
SCL-40 | 1 | |
LC-40B XR | 1 | |
DGU-403 | 1 | |
SIL-40C XR | ||
CTO-40C | 1 | |
SPD-M40 | 1 | |
CRB-40 | 1 | |
D582/F LCsolution Ver.1.25 AD | 1 | |
GLP 0.1φX2M | 1 | |
GLP 0.5φX1M | 1 | |
GLP 0.5φX5M | 1 | |
FEMALE T 1.6FT316L-0.4 | 1 | |
PEEK Male Nut, 5pcs/pack | 1 |
High Performance Liquid Chromatography
P/N | Description | Q'ty |
---|---|---|
LC-40D | 2 | |
SCL-40 | 1 | |
DGU-403 | 1 | |
SIL-40C | 1 | |
CTO-40C | 1 | |
ELSD-LTIII | 1 | |
Imtakt Presto FF-C18(150 mm×4.6 mmI.D.;2 μm) | 1 | |
Labtotal Vial KIT | 1 | |
MR20 μL | 1 | |
Piping kit 0.3 mm | 1 | |
Cable Assy | 1 | |
LC LabSolutions Single LC | 1 | |
REGULATOR,W/FILTER | 1 |
High Performance Liquid Chromatography
P/N | Description | Q'ty |
---|---|---|
CBM-40 | 1 | |
LC-40D | 2 | |
DGU-403 | 1 | |
SIL-40C | 1 | |
CTO-40C | 1 | |
CDD-10Avp | 1 | |
TUBING ASSY | 1 | |
SPD-40 | 1 | |
LabsolutionsLCGC,Single | 1 | |
RESERVOIR TRAY | 1 | |
WASHING PUMP SEAL KIT | 2 | |
Bottle, 250ml | 2 | |
Vial plate 1.5mL | 2 | |
COLUMN CALMP B8 | 1 | |
CABLE,HFBR3600-1-021 | 4 | |
FITTING SET 20A | 2 | |
TOOL KIT | 1 | |
Shim-pack SCR-102H | 2 | |
SCR-102(H) | 1 | |
Shim-pack GIST-HP C18-AQ | 1 |
High Performance Liquid Chromatography
P/N | Description | Q'ty |
---|---|---|
LC-40D | 2 | |
CBM-40Lite | 1 | |
DGU-403 | 1 | |
SIL-40C | 1 | |
CTO-40C | 1 | |
SPD-40V | 1 | |
RESERVOIR TRAY | 1 | |
Glass Reservoir Bottle, w/Cap, 5pcs | 1 | |
Mobile Phase Monitor (Contoller) | 1 | |
Mobile Phase Monitor(bottle holder) | 1 | |
Mobile Phase Monitor AC Adapter CN | 1 | |
CMD CABLE | 1 | |
MPM Bottle Holder Kit,Eeathquake-R | 1 | |
BELT | 1 | |
LOCKING PLATE | 1 | |
SYSTEM BASE KIT B | 1 | |
LabSolutions LCGC,DB Single LC | 1 | |
LC LabSolutions | 1 | |
SHIMADZU Vials, TORAST-H Glass Vial, 1.5mL | 1 | |
Shim-pack VP-Phenyl | 1 | |
Shim-pack Scepter C18-120 A | 1 |
High Performance Liquid Chromatography
P/N | Description | Q'ty |
---|---|---|
CBM-40 | 1 | |
LC-40D | 2 | |
WASHING PUMP SEAL KIT | 2 | |
Bottle, 250ml | 1 | |
DGU-403 | 1 | |
SIL-40C | 1 | |
Vial plate 1.5mL | 1 | |
CTO-40C | 1 | |
COLUMN CALMP B8 | 2 | |
CDD-10Avp | 1 | |
RESERVOIR TRAY | 1 | |
CABLE,HFBR3600-1-021 | 1 | |
TUBING ASSY | 1 | |
Shim-pack SCR-102(H) | 2 | |
GUARD COLUMN FOR SCR-102H | 1 | |
Reagent Kit for organic acid analysis system | 1 | |
LabSolutions LCGC,DB Single LC | 1 | |
FITTING SET 20A | 1 | |
PEEK Tube, 0.13mmIDx3000mm | 1 | |
PEEK Tube, 0.25mmIDx3000mm | 1 | |
PEEK Male Nut | 1 | |
CORD,UC-975-N01 | 1 | |
TOOL KIT | 1 | |
LC PC LabSolutions PC | 1 |
High Performance Liquid Chromatography
P/N | Description | Q'ty |
---|---|---|
SCL-40 | 1 | |
LC-40B XR | 1 | |
DGU-405 | 1 | |
SIL-40CXR | 1 | |
CTO-40S | 1 | |
SPD-M40 | 1 | |
MR 180μLⅡ | 1 | |
LC WORKSTATION | 1 | |
LC WORKSTATION Labsolution | 1 |
High Performance Liquid Chromatography
P/N | Description | Q'ty |
---|---|---|
SCL-40 | 1 | |
LC-40B XR | 1 | |
DGU-403 | 1 | |
SIL-40CXR | 1 | |
Vial plate 1.5mL | 1 | |
CTO-40S CN | 1 | |
SPD-M40 | 1 | |
MR 180 μL II | 1 | |
PIPING KIT B, for HPGE ID0.1 | 1 | |
Optical Fiber Cable Assy, Plastic, for Nexera Series,C | 1 | |
UC-975-N01 2.5m | 1 | |
LabSolutions LCGC,DB Multi LC-PDA | 1 |
High Performance Liquid Chromatography
P/N | Description | Q'ty |
---|---|---|
SCL-40 | 1 | |
LC-40D XR | 1 | |
DGU-40A5 | 1 | |
LPGE UNIT | 1 | |
SIL-40 | 1 | |
CTO-40C | 1 | |
SPD-M40A | 1 | |
LC WORKSTATION | 1 | |
LC WORKSTATION LabSolutions | 1 | |
SEMI MICRO TEMP CONT CELL | 1 | |
Mixer (300 uL) | 1 |
- Overview
- Features
- Applications
- Specs
- Downloads
- Configuration
- Columns
- Consumables
- Options
- Case Studies
- Support