EZ Test - Specs
Compact Tabletop Tester
Spec
Name | EZ Test | |||
---|---|---|---|---|
EZ-SX | EZ-LX | EZ-LX HS | ||
Tester Load Capacity (note 1) | Max. 500 N | Max. 5 kN | Max. 2 kN | |
The load cell type can be selected from 9 types: 1 N, 2 N, 5 N, 10 N, 20 N, 50 N, 100 N, 200 N, or 500 N. | The load cell type can be selected from 12 types: 1 N, 2 N, 5 N, 10 N, 20 N, 50 N, 100 N, 200 N, 500 N, 1 kN, 2 kN or 5 kN. (Up to 2 kN for EZ-LX HS) | |||
Load Method | High-precision constant-speed strain measurement using backlash-free ball screw drive | |||
Test Force Measurement Load | High-Precision Type (note 2) | ±0.5% of indicated value (within 1/500 to 1/1 of load cell rated capacity) | ||
Conforms to JIS B 7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4. | ||||
Standard-Precision Type (note 2) | ±1% of indicated value (within 1/500 to 1/1 of load cell rated capacity) | |||
Conforms to JIS B 7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4. | ||||
Range | 1 range (rangeless) | |||
Test Force Calibration | Automatic calibration using calibration cable | |||
Crosshead Speed Range | 0.001 to 1000 mm/min | 0.001 to 2000 mm/min | ||
Maximum Return Speed | 1500 mm/min | 3000 mm/min | ||
Crosshead Speed Accuracy (Note 3) | Within ±0.1% of test speed | |||
Crosshead Speed and Allowable Test Force | Up to the capacity of the load cell at all speeds | |||
Distance Between Crosshead and Jig Mounting Surface | 500 mm | 920 mm | ||
Maximum Grip Space | 395 mm (500 N max. load cell + tensile jig) | 700 mm (5 kN load cell + 5 kN screw type flat grips) | ||
755 mm (1 kN load cell + 1 kN screw type flat grips) | ||||
860 mm (500 N max. load cell + tensile jig) | ||||
Depth of Test Space | 100 mm (table section) | |||
Crosshead Position Detection | Measurement & Display | Optical encoder measurement, digital display (display resolution: 1 µm) | ||
Accuracy | ±0.1% of indicated value or ±0.01 mm, whichever is greater | |||
Crosshead Control | Single test control (single-direction tension or compression test), cycle test control (repetitive tension or compression test) | |||
Sampling Speed | 1 ms max. (TRAPEZIUMX-V/TRAPEZIUM LITE X is needed for this function) | |||
Standard Functions Included | Constant test force (creep) control (note 4) | |||
Auto-stop and auto-return functions when specimen fracture is detected (crosshead auto home-position return) | ||||
Test condition le function, adjustable crosshead speed function | ||||
Display function: Actual test force display or stress display (set by user) | ||||
Crosshead displacement display in mm or %/GL (set by user) | ||||
Peak point test force and stroke | ||||
Test force and displacement analog output: 0 V to 5 V DC full scale, respectively (for external recorder) | ||||
USB interface | ||||
Manual crosshead position fine adjustment | ||||
Adjustable controller | ||||
Touch load alarm | ||||
Dimensions and Weight | W400 × D530 × H885 mm, approx. 33 kg | W400 × D530 × H1315 mm, approx. 55 kg | ||
Input Power Supply Voltage (Note 5) | Single phase, 100 V to 150 V AC, 50/60 Hz, or 200V to 230V AC, 50/60 Hz | |||
Power Capacity | 700 VA | 850 VA | ||
Installation Environmental Conditions | Temperature: 5℃ to 40℃, Humidity: 20% to 80% (no condensation) | |||
Power voltage fluctuation: within ±10%, Vibration: 10 Hz max., Amplitude: 5 µm max. |
Note 1: When the load cell capacity is smaller than the tester load capacity, the former is the maximum test force.
Note 2: Shimadzu recommends inspection at an installation site that meets the requirements specified in these standards.
Note 3: Crosshead speed accuracy is calculated from the crosshead displacement within a prescribed time at a constant speed between 0.5 mm/min and 500 mm/min.
Note 4: The test force is kept constant at 70% or less of the tester load capacity, for up to 12 hours.
Note 5: Ground resistance should be 100 Ω or less.
EZ-SX
Tester Load Capacity (Note 1) |
Max. 500 N |
---|---|
Crosshead Speed Range |
0.001 to 1000 mm/min |
Maximum Return Speed |
1500 mm/min |
Test Force Measurement : High-Precision Type(Note 2) |
±0.5 % of indicated value (within 1/500 to 1/1 of load cell rated capacity) |
Test Force Measurement : Standard-Precision Type(Note 2) |
±1 % of indicated value (within 1/500 to 1/1 of load cell rated capacity) |
Test Force Measurement : Range |
1 range (rangeless) |
Test Force Measurement : Test Force Calibration |
Automatic calibration using calibration cable |
Dimensions |
W400 × D530 × H885 mm |
Weight |
Approx. 33 kg |
Note 1: When the load cell capacity is smaller than the tester load capacity, the former is the maximum test force.
Note 2: Shimadzu recommends validation at an installation site that meets the requirements specified in these standards.
EZ-LX
Tester Load Capacity(Note 1) |
Max. 5 kN |
---|---|
Crosshead Speed Range |
0.001 to 1000 mm/min |
Maximum Return Speed |
1500 mm/min |
Test Force Measurement : High-Precision Type(Note 2) |
±0.5 % of indicated value (within 1/500 to 1/1 of load cell rated capacity) |
Test Force Measurement : Standard-Precision Type(Note 2) |
±1 % of indicated value (within 1/500 to 1/1 of load cell rated capacity) |
Test Force Measurement : Range |
1 range (rangeless) |
Test Force Measurement : Test Force Calibration |
Automatic calibration using calibration cable |
Dimensions |
W400 × D530 × H1315 mm |
Weight |
Approx. 55 kg |
Note 1: When the load cell capacity is smaller than the tester load capacity, the former is the maximum test force.
Note 2: Shimadzu recommends validation at an installation site that meets the requirements specified in these standards.
EZ-LX HS
Tester Load Capacity(Note 1) |
Max. 2 kN |
---|---|
Crosshead Speed Range |
0.001 to 2000 mm/min |
Maximum Return Speed |
3000 mm/min |
Test Force Measurement : High-Precision Type(Note 2) |
±0.5 % of indicated value (within 1/500 to 1/1 of load cell rated capacity) |
Test Force Measurement : Standard-Precision Type(Note 2) |
±1 % of indicated value (within 1/500 to 1/1 of load cell rated capacity) |
Test Force Measurement : Range |
1 range (rangeless) |
Test Force Measurement : Test Force Calibration |
Automatic calibration using calibration cable |
Dimensions |
W400 × D530 × H1315 mm |
Weight |
Approx. 55 kg |
Note 1: When the load cell capacity is smaller than the tester load capacity, the former is the maximum test force.
Note 2: Shimadzu recommends validation at an installation site that meets the requirements specified in these standards.
EZ-SX
Load capacity |
500 N |
---|---|
Test speed |
0.001 to 1,000 mm/mim |
Return speed |
1,500 mm/min |
Test force measurement accuracy : Highprecision type |
±0.5 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4 |
Test force measurement accuracy : Standard precision type |
±1 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4 |
EZ-LX
Load capacity |
5 kN |
---|---|
Test speed |
0.001 to 1,000 mm/mim |
Return speed |
1,500 mm/min |
Test force measurement accuracy : Highprecision type |
±0.5 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4 |
Test force measurement accuracy : Standard precision type |
±1 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4 |
EZ-LX HS
Load capacity |
2 kN |
---|---|
Test speed |
0.001 to 2,000 mm/min |
Return speed |
3,000 mm/min |
Test force measurement accuracy : Highprecision type |
±0.5 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4 |
Test force measurement accuracy : Standard precision type |
±1 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4 |