MCT Series - Specs
Micro Compression Tester
510
Loading Unit : Loading Method |
Electromagnetic force |
---|---|
Loading unit : Loading range (mN) |
9.8 – 4903 |
Loading unit : Load Accuracy |
Within ± 1% of displayed test force or 0.1 mN (whichever is greater) |
Loading unit : Resolution (µm) |
5 (when testing at 49 mN or less) |
Displacement measurement unit : Method |
Differential transformer |
Displacement measurement unit : Measurement range (µm) |
0 – 100 |
Displacement measurement unit : Min. increment (µm) |
0.001 |
Displacement measurement unit : Linearity |
Within ± 2% of full scale |
Optical monitor : Total magnification |
Approx. × 100, × 500 (× 200, × 400, × 1000 with option) |
Optical Monitor : Objective Lens |
× 10, × 50 (× 20, × 40, × 100 with option) |
Optical Monitor : Eyepiece |
× 10 |
Optical monitor : Illumination method |
Epiluminescent |
Optical monitor : Illumination lamp |
LED 3 W / 3 V |
Optical monitor : Light path |
Switching between observation and photography possible |
Optical Head : Collimation method |
Direct connection between encorder and control handle: synchronized movement of two indexes |
Optical Head : Detector |
Optical encoder |
Optical Head : Effective measurement range |
Approx. 200 µm (with × 50 objective lens) |
Optical Head : Min. increment |
0.01 µm / pulse |
Indenter : Upper pressure indenter |
Type: Flat indenter (50 µm diameter) |
Indenter : Lower pressure plate |
SKS flat plate (Diamond pressure plate available as option) |
Specimen stage : Vertical positioning range |
Approx. 60 mm |
Specimen Stage : Area |
Approx. W 125 mm × D 125 mm |
Specimen stage : Horizontal positioning range |
25 mm for both X and Y directions |
Available test parameters : Test mode |
Compression test, load-unload test, cyclic test |
Available test parameters : Shape of specimen |
Particle, fiber or other shapes |
Available test parameters : Data Processing Items |
Calculation of compressive strength, display of test parameters and results |
Available test parameters : Power Supply |
Single phase AC 100–115 V ± 10%, AC 230 V ± 10% (Ground resistance 100 Ω max.) |
Utilities : Power Consumption |
Approx. 100 W (not including power consumption of PC) |
Utilities : Grounding*1 |
The grounding pins of 3-pin connectors must be connected to a resistance of 100 W or less. |
Utilities : Temperature |
Recommended temperature: 23 ± 1°C Allowable range: 10°C to 35°C |
Utilities : Vibration |
Horizontal vibration: 0.017 Gal max. (at 10 Hz or more) |
Utilities : Humidity |
80% max. (no condensation) |
External Dimensions |
Tester: Approx. W 355 × D 405 × H 530 mm Control unit: Approx. W 315 × D 375 × H 110 mm |
Weight |
Tester: Approx. 60 kg Control unit: Approx. 5 kg |
*1 This product is a precision measuring device. Special consideration is required for the product's installation conditions.
511
Loading Unit : Loading Method |
Electromagnetic force |
---|---|
Loading unit : Loading range (mN) |
9.8 – 4903 |
Loading unit : Load Accuracy |
Within ± 1% of displayed test force or 0.1 mN (whichever is greater) |
Loading unit : Resolution (µm) |
5 (when testing at 49 mN or less) |
Displacement measurement unit : Method |
Differential transformer |
Displacement measurement unit : Measurement range (µm) |
0 – 10 |
Displacement measurement unit : Min. increment (µm) |
0.0001 |
Displacement measurement unit : Linearity |
Within ± 2% of full scale |
Optical monitor : Total magnification |
Approx. × 100, × 500 (× 200, × 400, × 1000 with option) |
Optical Monitor : Objective Lens |
× 10, × 50 (× 20, × 40, × 100 with option) |
Optical Monitor : Eyepiece |
× 10 |
Optical monitor : Illumination method |
Epiluminescent |
Optical monitor : Illumination lamp |
LED 3 W / 3 V |
Optical monitor : Light path |
Switching between observation and photography possible |
Optical Head : Collimation method |
Direct connection between encorder and control handle: synchronized movement of two indexes |
Optical Head : Detector |
Optical encoder |
Optical Head : Effective measurement range |
Approx. 200 µm (with × 50 objective lens) |
Optical Head : Min. increment |
0.01 µm / pulse |
Indenter : Upper pressure indenter |
Type: Flat indenter (50 µm diameter) |
Indenter : Lower pressure plate |
SKS flat plate (Diamond pressure plate available as option) |
Specimen stage : Vertical positioning range |
Approx. 60 mm |
Specimen Stage : Area |
Approx. W 125 mm × D 125 mm |
Specimen stage : Horizontal positioning range |
25 mm for both X and Y directions |
Available test parameters : Test mode |
Compression test, load-unload test, cyclic test |
Available test parameters : Shape of specimen |
Particle, fiber or other shapes |
Available test parameters : Data Processing Items |
Calculation of compressive strength, display of test parameters and results |
Available test parameters : Power Supply |
Single phase AC 100–115 V ± 10%, AC 230 V ± 10% (Ground resistance 100 Ω max.) |
Utilities : Power Consumption |
Approx. 100 W (not including power consumption of PC) |
Utilities : Grounding*1 |
The grounding pins of 3-pin connectors must be connected to a resistance of 100 W or less. |
Utilities : Temperature |
Recommended temperature: 23 ± 1°C Allowable range: 10°C to 35°C |
Utilities : Vibration |
Horizontal vibration: 0.017 Gal max. (at 10 Hz or more) |
Utilities : Humidity |
80% max. (no condensation) |
External Dimensions |
Tester: Approx. W 355 × D 405 × H 530 mm Control unit: Approx. W 315 × D 375 × H 110 mm |
Weight |
Tester: Approx. 60 kg Control unit: Approx. 5 kg |
*1 This product is a precision measuring device. Special consideration is required for the product's installation conditions.
210
Loading Unit : Loading Method |
Electromagnetic force |
---|---|
Loading unit : Loading range (mN) |
9.8 – 1961 |
Loading unit : Load Accuracy |
Within ± 1% of displayed test force or 0.1 mN (whichever is greater) |
Loading unit : Resolution (µm) |
2 (when testing at 19 mN or less) |
Displacement measurement unit : Method |
Differential transformer |
Displacement measurement unit : Measurement range (µm) |
0 – 100 |
Displacement measurement unit : Min. increment (µm) |
0.001 |
Displacement measurement unit : Linearity |
Within ± 2% of full scale |
Optical monitor : Total magnification |
Approx. × 100, × 500 (× 200, × 400, × 1000 with option) |
Optical Monitor : Objective Lens |
× 10, × 50 (× 20, × 40, × 100 with option) |
Optical Monitor : Eyepiece |
× 10 |
Optical monitor : Illumination method |
Epiluminescent |
Optical monitor : Illumination lamp |
LED 3 W / 3 V |
Optical monitor : Light path |
Switching between observation and photography possible |
Optical Head : Collimation method |
Direct connection between encorder and control handle: synchronized movement of two indexes |
Optical Head : Detector |
Optical encoder |
Optical Head : Effective measurement range |
Approx. 200 µm (with × 50 objective lens) |
Optical Head : Min. increment |
0.01 µm / pulse |
Indenter : Upper pressure indenter |
Type: Flat indenter (50 µm diameter) |
Indenter : Lower pressure plate |
SKS flat plate (Diamond pressure plate available as option) |
Specimen stage : Vertical positioning range |
Approx. 60 mm |
Specimen Stage : Area |
Approx. W 125 mm × D 125 mm |
Specimen stage : Horizontal positioning range |
25 mm for both X and Y directions |
Available test parameters : Test mode |
Compression test, load-unload test, cyclic test |
Available test parameters : Shape of specimen |
Particle, fiber or other shapes |
Available test parameters : Data Processing Items |
Calculation of compressive strength, display of test parameters and results |
Available test parameters : Power Supply |
Single phase AC 100–115 V ± 10%, AC 230 V ± 10% (Ground resistance 100 Ω max.) |
Utilities : Power Consumption |
Approx. 100 W (not including power consumption of PC) |
Utilities : Grounding*1 |
The grounding pins of 3-pin connectors must be connected to a resistance of 100 W or less. |
Utilities : Temperature |
Recommended temperature: 23 ± 1°C Allowable range: 10°C to 35°C |
Utilities : Vibration |
Horizontal vibration: 0.017 Gal max. (at 10 Hz or more) |
Utilities : Humidity |
80% max. (no condensation) |
External Dimensions |
Tester: Approx. W 355 × D 405 × H 530 mm Control unit: Approx. W 315 × D 375 × H 110 mm |
Weight |
Tester: Approx. 60 kg Control unit: Approx. 5 kg |
*1 This product is a precision measuring device. Special consideration is required for the product's installation conditions.
211
Loading Unit : Loading Method |
Electromagnetic force |
---|---|
Loading unit : Loading range (mN) |
9.8 – 1961 |
Loading unit : Load Accuracy |
Within ± 1% of displayed test force or 0.1 mN (whichever is greater) |
Loading unit : Resolution (µm) |
2 (when testing at 19 mN or less) |
Displacement measurement unit : Method |
Differential transformer |
Displacement measurement unit : Measurement range (µm) |
0 – 10 |
Displacement measurement unit : Min. increment (µm) |
0.0001 |
Displacement measurement unit : Linearity |
Within ± 2% of full scale |
Optical monitor : Total magnification |
Approx. × 100, × 500 (× 200, × 400, × 1000 with option) |
Optical Monitor : Objective Lens |
× 10, × 50 (× 20, × 40, × 100 with option) |
Optical Monitor : Eyepiece |
× 10 |
Optical monitor : Illumination method |
Epiluminescent |
Optical monitor : Illumination lamp |
LED 3 W / 3 V |
Optical monitor : Light path |
Switching between observation and photography possible |
Optical Head : Collimation method |
Direct connection between encorder and control handle: synchronized movement of two indexes |
Optical Head : Detector |
Optical encoder |
Optical Head : Effective measurement range |
Approx. 200 µm (with × 50 objective lens) |
Optical Head : Min. increment |
0.01 µm / pulse |
Indenter : Upper pressure indenter |
Type: Flat indenter (50 µm diameter) |
Indenter : Lower pressure plate |
SKS flat plate (Diamond pressure plate available as option) |
Specimen stage : Vertical positioning range |
Approx. 60 mm |
Specimen Stage : Area |
Approx. W 125 mm × D 125 mm |
Specimen stage : Horizontal positioning range |
25 mm for both X and Y directions |
Available test parameters : Test mode |
Compression test, load-unload test, cyclic test |
Available test parameters : Shape of specimen |
Particle, fiber or other shapes |
Available test parameters : Data Processing Items |
Calculation of compressive strength, display of test parameters and results |
Available test parameters : Power Supply |
Single phase AC 100–115 V ± 10%, AC 230 V ± 10% (Ground resistance 100 Ω max.) |
Utilities : Power Consumption |
Approx. 100 W (not including power consumption of PC) |
Utilities : Grounding*1 |
The grounding pins of 3-pin connectors must be connected to a resistance of 100 W or less. |
Utilities : Temperature |
Recommended temperature: 23 ± 1°C Allowable range: 10°C to 35°C |
Utilities : Vibration |
Horizontal vibration: 0.017 Gal max. (at 10 Hz or more) |
Utilities : Humidity |
80% max. (no condensation) |
External Dimensions |
Tester: Approx. W 355 × D 405 × H 530 mm Control unit: Approx. W 315 × D 375 × H 110 mm |
Weight |
Tester: Approx. 60 kg Control unit: Approx. 5 kg |
*1 This product is a precision measuring device. Special consideration is required for the product's installation conditions.