Microservo MMT Series - Specs

Magnetic Micro Testing System

Add Bookmark

MMT-500NV-10

Test Force

±500N

Piston stroke

±10mm

Cycle speed

100Hz

Controlled items

Test force and stroke (two can be added as options)

Indication accuracy : Test force

Within ±0.5 % of indicated value or ±0.02 % of maximum dynamic test force, whichever is greater

Indication accuracy : Stroke

Within +1 % of indicated value or ±0.1 % of maximum stroke, whichever is greater

Installation space (W × D × H)

Approx. 1000 × 500 × 1200 mm

Actuator mount

Either top or bottom mount

Total weight

Approx. 150 kg

Power Requirements

1Ø 100V 1kVA

Site requirements

- Minimal temperature variations (+10 to +40 ℃ recommended, with temperature variations within ±5 ℃)
- Low humidity
- Not exposed to direct air flow from heating or cooling systems
- No direct sunlight
- Low dust
- No significant vibration

Read more
Show Less

MMT-250NV-10

Test Force

±250N

Piston stroke

±10mm

Cycle speed

100Hz

Controlled items

Test force and stroke (two can be added as options)

Indication accuracy : Test force

Within ±0.5 % of indicated value or ±0.02 % of maximum dynamic test force, whichever is greater

Indication accuracy : Stroke

Within +1 % of indicated value or ±0.1 % of maximum stroke, whichever is greater

Installation space (W × D × H)

Approx. 1000 × 500 × 1200 mm

Actuator mount

Either top or bottom mount

Total weight

Approx. 120 kg

Power Requirements

1Ø 100V 1kVA

Site requirements

- Minimal temperature variations (+10 to +40 ℃ recommended, with temperature variations within ±5 ℃)
- Low humidity
- Not exposed to direct air flow from heating or cooling systems
- No direct sunlight
- Low dust
- No significant vibration

Read more
Show Less

Test Force

±100N

Piston stroke

±10mm

Cycle speed

100Hz

Controlled items

Test force and stroke (two can be added as options)

Indication accuracy : Test force

Within ±0.5 % of indicated value or ±0.02 % of maximum dynamic test force, whichever is greater

Indication accuracy : Stroke

Within +1 % of indicated value or ±0.1 % of maximum stroke, whichever is greater

Installation space (W × D × H)

Approx. 1000 × 500 × 1200 mm

Actuator mount

Either top or bottom mount

Total weight

Approx. 100 kg

Power Requirements

1Ø 100V 1kVA

Site requirements

- Minimal temperature variations (+10 to +40 ℃ recommended, with temperature variations within ±5 ℃)
- Low humidity
- Not exposed to direct air flow from heating or cooling systems
- No direct sunlight
- Low dust
- No significant vibration

Read more
Show Less

MMT-101NV-2

Test Force

±100N

Piston stroke

±2mm

Cycle speed

100Hz

Controlled items

Test force and stroke (two can be added as options)

Indication accuracy : Test force

Within ±0.5 % of indicated value or ±0.02 % of maximum dynamic test force, whichever is greater

Indication accuracy : Stroke

Within +1 % of indicated value or ±0.1 % of maximum stroke, whichever is greater

Installation space (W × D × H)

Approx. 1000 × 500 × 1200 mm

Actuator mount

Either top or bottom mount

Total weight

Approx. 100 kg

Power Requirements

1Ø 100V 1kVA

Site requirements

- Minimal temperature variations (+10 to +40 ℃ recommended, with temperature variations within ±5 ℃)
- Low humidity
- Not exposed to direct air flow from heating or cooling systems
- No direct sunlight
- Low dust
- No significant vibration

Read more
Show Less

MMT-11NV-2

Test Force

±10N

Piston stroke

±2mm

Cycle speed

60Hz

Controlled items

Test force and stroke (two can be added as options)

Indication accuracy : Test force

Within ±1 % of indicated value or ±0.02 % of maximum dynamic test force, whichever is greater

Indication accuracy : Stroke

Within +1 % of indicated value or ±0.1 % of maximum stroke, whichever is greater

Installation space (W × D × H)

Approx. 1000 × 500 × 1200 mm

Actuator mount

Bottom

Total weight

Approx. 80 kg

Power Requirements

1Ø 100V 500VA

Site requirements

- Minimal temperature variations (+10 to +40 ℃ recommended, with temperature variations within ±5 ℃)
- Low humidity
- Not exposed to direct air flow from heating or cooling systems
- No direct sunlight
- Low dust
- No significant vibration

Read more
Show Less

MMT11NV-2

Test force capacity

±10 N

Stroke

±2 mm

Max. frequency

60 Hz

Power Requirements

AC100 V

Read more
Show Less

-101NV-10

Test force capacity

±100 N

Stroke

±10 mm

Max. frequency

60 Hz

Power Requirements

AC100 V

Read more
Show Less

-250NV-10

Test force capacity

±250 N

Stroke

±10 mm

Max. frequency

60 Hz

Power Requirements

AC100 V

Read more
Show Less
{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3469,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3469,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}