4830 Controller - Specs

Servo Controller for Fatigue and Endurance System

Add Bookmark

Servo Controller 4830

Display unit

5.7-inch color LCD

Control panel

Touch panel, function keys, jog dial, numeric keypad, test operation keys, power unit operation keys

Test waveforms

Sine, triangular, rectangular, haversine, haver-triangular, trapezoidal, ramp, 1/2 haversine, step, sweep, and random waves, external input, programmed waves (optional*1), file waves optional*1)

Test frequency

0.00001 to 1000 Hz

Slow settings

Slow start/stop

Test parameter registration

Max. 9 parameters

Waveform display functions

Time, X-Y, and peak waveforms

Measurement functions

1 range (rangeless) 24-bit
Max. 40 kHz sampling with 4 acquisition channels
Linear correction (linearization) function

Size

W350 × D420 × H148 mm

Control method

Full digital two-degree-of-freedom PID

Control functions

Amplitude, average gain correction (AGC), PID autotuning, sample anti-overloading function (contact load), user-specified phase differential control by synchronized operation, waveform distortion correction*1(transfer function correction)

Limit functions

Measurement value 4-point limiter, cycle counter, external input

Communications functions

USB interface

Other functions

Calculation function (such as adding, subtracting, averaging, and stress/strain), push testing function, consumable consumption time management function

External input/output

Analog Output: 4 channels (±10 V), Input: 1 channel (±10 V)
For monitoring or waveform input
Digital Output: 8 channels Input: 8 channels

Control signal input

Test force (TD1), stroke (TD2), and external input (AUX)
Note: Up to two amplifiers can be added as an option.

Power Requirements

Single-phase AC 100 to 230 V*2 50/60 Hz 300 VA

*1 Only during software use
*2 The standard power cord included with the system is only for AC 100 V.

Read more
Show Less

4830 Controller

Test waveforms

Sine, triangular, rectangular, etc.

Amplifier

Test force, stroke

Control mode

Test force, stroke, virtual transducer

Power Requirements

Single-phase 100 V, 300 VA

Read more
Show Less
{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3480,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3480,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}