EDX-8100 - Specs

Energy Dispersive X-ray Fluorescence Spectrometer

Add Bookmark

EDX-7000/8000/8100

Measurement principle

X-ray fluorescence spectrometry

Measurement method

Energy dispersion

Target samples

Solids, liquids, powders

Measuring range

6C to 92U

Sample size

W 300 x D 275 x approx.H 100 mm (excluding radiuses)

Maximum sample mass

5kg (200g per sample when using turret, Gross mass 2.4kg)

Dose rate

1 µSv/h or less.

Read more
Show Less

X-ray generator

X-ray tube

Rh target (Standard model/Premium model)*1

Voltage

4 kV to 50 kV

Current

1 µA to 1000 µA

Cooling method

Air-cooled (with fan)

Irradiated area

Automatic switching in four stages: 1, 3, 5, and 10 mm diameter
Automatic switching in four stages: 0.3, 1, 3, and 10 mm diameter*2

Primary filters

Five types (six, including the open position), automatic replacement

*1 Option for EDX-8100
*2 Microsoft® Office is not included.

Read more
Show Less

Detector

Type

Silicon drift detector (SDD)

Liquid nitrogen

Not required (electronic cooling)

Sample chamber

Measurement atmosphere

Air, vacuum*1, helium (He)*2

Sample replacement*

12-sample turret

Sample observations

Semiconductor camera

*1 Option for EDX-8100
*2 Option for EDX-7000

Data processor

CPU

Intel Core i5 or above

Memory

4 GB min.

HDD

250 GB min.

Optical drive

Super multi drive

OS

Windows™ 10 Pro (64-bit)*

* Microsoft Office is not included.

Read more
Show Less

Software

Qualitative analysis

Measurement/analysis software

Quantitative analysis

Calibration curve method, correction for coexistent elements, FP method,film FP method, background FP method

Matching software

Intensity/content

Utilities

Automatic calibration functions (energy calibration, FWHM calibration)

Others

Instrument status monitoring function,
analysis results tabulation function

Read more
Show Less

Installation

Temperature

10 °C to 30 °C (temperature fluctuation rate 2 °C/hour max.,
temperature fluctuation range: 10 °C max.)

Relative humidity

40 % to 70 % (no condensation)

Power supply

100-240 V AC ±10 %, 2 A earthed socket

Dimensions

W 460 x D 590 x H 360 mm

Weight

Approx. 45 kg

Options

Vacuum measurement unit, helium purge unit, turret unit, screening analysis kits

Read more
Show Less
{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3380,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3380,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}